NEPS 1000 Leak Testing

Leak testing can be carried out to ensure the suitability of the equipment for pressurized purging. The default pressure test setting is 2.5 psi.

Leak testing with the NEPS nitrogen purge system is a simple operation and has a sensitivity which is related to pressure degradation against time. The NEPS nitrogen purge system has a display resolution of 0.01 psi allowing users to achieve high sensitivities depending on the monitored time for the equipment under test to degrade from the initial test pressure.

The equipment volume must be considered when calculating actual leak rate.

Typical sensitivities for NEPS 1000 are:

NEPS 1000 Leak Testing Pressure Degradation Chart

60 Minutes to degrade by 0.01 psi

1,000 cc volume = 1.9 x 10-4 cc/second

10,000 cc volume = 1.9 cc x 10-3 cc/second

Caution:

Leak testing must be conducted with a stable/constant ambient/equipment temperature. A variation of 3.6°F (2°C) can cause up to a 0.1 psi change in pressure.