NEPS 1000 Leak Testing

 

Leak testing can be carried out at three pressure levels depending on the mechanical strength of the equipment to be tested.

The three pressure levels are:        
1.5 psi (10.3 Kpa)
2.5 psi (17.2 Kpa)
5.0 psi (34.4 Kpa)
  

Leak testing with the NEPS nitrogen purge system is a simple operation and has a sensitivity which is related to pressure degradation against time.  The NEPS nitrogen purge system has a display resolution of 0.01 psi allowing users to achieve high sensitivities depending on the monitored time for the equipment under test to degrade from the initial test pressure.

The rate of leakage must take into consideration the equipment volume. 

Pressure Degradation GraphTypical Sensitivities for NEPS 1000 are:

60 Minutes to degrade by 0.01 psi
1,000 cc volume = 1.9 x 10-4 cc/second
10,000 cc volume = 1.9 cc x 10-3 cc/second

Caution: Leak testing must be conducted with a stable/constant ambient/equipment temperature.  A variation of 3.6°F (2°C) can cause up to a 0.1 psi change in pressure.

   
   